Dry heat is used for sterilization instead of the much more efficient moist heat because some materials are sensitive to moisture. Dry heat is often used to ensure that glass and other laboratory equipment is free of pyrogenic material. The process of sterilization within a chamber or hot air tunnel is a critical process and there is a regulatory requirement for validation of the process in most countries. Validation is defined as the documented procedure of obtaining, recording and interpreting results to ensure that the dry heat sterilization process has been and will be consistently effective. Dry heat sterilizer validation consists of accurately measuring the temperature at critical points within the sterilization chamber throughout the process. Dry heat process generally employs a temperature between 250°C and 400°C for varying time. The sterilizer is required to heat all parts of its load up to the specified temperature for a specified period long enough to achieve the desired sterility.
Smoke flow direction with the Tunnel Sterilizer |
Temperature Mapping: During the mapping period, the temperature readings for the temperature recording chart and the average chamber temperature for Tunnels at 250*C and greater are consistently ± 15.0*C of each other throughout the study:Temperature range among thermocouples for Tunnels at 250*C and greater, for each five minute interval, is ± 15.0*C of the average temperature throughout the study. The average temperature for Tunnels at 250*C and greater at each interval is ± 15.0*C of setpoint temperatureCalculation of tunnel speed and cycle timing is a important issue during validation of Sterilization and Depyrogenation tunnel. We can solve with a easy two steps mention below.
Example for Conveyor speed calculations,
# Vial Dimension 50 mm
# Speed of conveyor 135 mm/min
# Length of the tunnel 457 mm
# Width of the tunnel 1200mm
No’s of vial lying vertically = 1200 mm/ 50 mm = 24 vials
No of vial lying horizontally = 457 mm / 50 mm = 09 vials
Total vial capacity in the tunnel sterilization zone = 24×09 = 216 vial
If 50 mm moved 09 vial will released from hot zone. In case of machine speed 135mm/min means
(9×135)/50 = 24 vials will come from hot zone/per min
Time spend by the each vials id 135 mm/50 mm = 2.7 min
What must be the Differential pressure across each zones of Depyrogenation tunnel. ??