BS EN 285 · ANNEX C · THERMOMETRIC TEST SIMULATOR
THERMOTEST
PRO
VIDEO TUTORIAL
A complete walkthrough of every feature — from the live temperature chart to automatic PASS/FAIL evaluation against EN 285.
By Palash Chandra Das  ·  pres.net.in
01 — OVERVIEW
What Is
ThermoTest Pro?
An interactive browser-based simulator that digitises the BS EN 285 small load thermometric test — Annex C — in real time. No installation. No spreadsheets.
Animates all three sensor traces simultaneously
Evaluates every EN 285 tolerance automatically
Delivers a clear PASS or FAIL verdict
Runs entirely in any modern web browser
// THERMOTEST PRO · EN 285
◉ PASS
STERIL. TEMP
134.0
°C (Ts)
SENSOR S₁
134.18
°C REF PT
SENSOR S₂
134.06
°C COLDEST
SENSOR S₃
134.24
°C +50mm
PASS
Equil: 10s / 15s ✔ | T₁: 0.12°C ✔ | T₂: 0.74°C ✔ | T₃: 0.18°C ✔ | Tᴮ: OK ✔
02 — TEMPERATURE CHART
Three Sensors.
One Chart.
A live animated reproduction of Figure C.1 from BS EN 285 Annex C.
S₁ Cyan — Reference point (drain). Reaches Ts first.
S₂ Orange — Coldest point in pack. Lags during equilibration.
S₃ Purple — 50 mm above pack. May overshoot initially.
Tᴮ Band — Dashed lines at Ts and Ts+3°C define the compliance window.
Point A = plateau start  ·  Point B = plateau end
A Ts Ts+3 150° t₂ t₄ holding
03 — CONTROLS
Running the
Simulation
Three controls give you complete playback control over the sterilization cycle.
▶ RUN — Starts the live animation. Traces draw in real time.
⏸ PAUSE — Freezes the chart at any moment for inspection.
↺ RESET — Clears all data and returns to the start.
Timeline Slider — Drag to scrub to any point. All readings update instantly.
TIMELINE — DRAG TO SCRUB
0s
420s t = 273s
⏸ PAUSE
↺ RESET
PHASE STATUS
HEAT-UP
EQ
HOLDING ●
COOL
EQUIL. TIME (t₂)
10.0
s (MAX 15s)
HOLDING TIME (t₄)
163.0
s ELAPSED
04 — ZOOM FEATURE
Full Cycle vs
t₂ Focus
Two precision views with a smooth animated transition between them.
⊟ Full Cycle — Complete view from cold start to cooling. Y-axis: 100°C to Ts+12°C.
⊕ t₂ Focus — X-axis centres on the equilibration window. Y-axis narrows to Ts band ± 2°C.
Minimap — Appears in Focus mode. Shows full-cycle context with zoom region in amber.
Crosshair Tooltip — Hover anywhere to read exact S₁, S₂, S₃ values at any time point.
⊟ FULL CYCLE VIEW
Ts +3
⊕ t₂ FOCUS VIEW — ACTIVE
Ts +3 t₂ window
⊕ FOCUS MODE: Transition is smoothly animated via live axis interpolation — you can see each axis slide to its new range.
05 — TOLERANCES
Live Tolerance
Monitoring
Every EN 285 tolerance is evaluated continuously during the simulation and shown as colour-coded live badges.
T₁ ≤ 2°C — Max Δ between S₁ and S₂ during the entire holding phase.
T₂ ≤ 5°C — Max Δ between S₁ and S₃ within the first 60 seconds of the plateau.
T₃ ≤ 2°C — Same S₁ vs S₃ difference, but after the first 60 seconds.
Tᴮ — All sensors within Ts → Ts+3°C. Checked only after t₂ elapses.
TOLERANCES — EN 285 · EVALUATED DURING HOLD
T₁Δ(S₁,S₂) max during holding0.12°C ✔
T₂Δ(S₁,S₃) first 60s (≤5°C)0.74°C ✔
T₃Δ(S₁,S₃) after 60s (≤2°C)0.18°C ✔
TᴮBand Ts → Ts+3°C after t₂OK ✔
FAILURE EXAMPLE — T₁ VIOLATED
T₁Δ(S₁,S₂) max during holding 3.47°C ✘
06 — CONFIGURATION
Customise Your
Test Parameters
The Config tab gives full control. Sliders update the simulation instantly on reset.
Sterilization Temp — 121°C to 135°C
Chamber Volume — Auto-switches limit: ≤800L→15s, >800L→30s
Equilibration Time — Set intentionally above limit to test FAIL mode
Noise Level — Simulates real-world sensor variability
Failure Mode — Add S₂/S₃ offsets to deliberately violate tolerances
STERILIZER SETUP
STERIL. TEMP
134°C
CHAMBER VOL
400 L
EQUIL. TIME
10 s
HOLD TIME
300 s
LOAD MASS
6 kg
NOISE LEVEL
1.0°
⚠ FAILURE MODE SIMULATION
S₂ offset (60s)
+2.5°C
07 — RESULT
Automatic
PASS / FAIL
Once the cycle completes, every tolerance is evaluated simultaneously. All must pass for the overall result to be PASS.
Equilibration time within the EN 285 volume limit
T₁ max delta ≤ 2°C throughout holding
T₂ max delta ≤ 5°C in first 60 seconds
T₃ max delta ≤ 2°C after 60 seconds
Tᴮ band maintained after t₂ throughout holding
PASS
Equil: 10s / 15s ✔ | T₁: 0.12°C ✔ | T₂: 0.74°C ✔ | T₃: 0.18°C ✔ | Tᴮ: OK ✔
FAIL
Equil: 18s / 15s ✘ | T₁: 3.47°C ✘ | T₂: 0.74°C ✔ | T₃: 0.18°C ✔ | Tᴮ: VIOLATED ✘
The status indicator at the top of the screen mirrors the result in real time — cycling from ◉ STANDBY◉ RUNNING◉ PASS or ◉ FAIL.
QUICK START GUIDE
Get Running in
60 Seconds
01
Open the Link
No install needed. Runs in any modern browser.
02
Config Tab
Set your Ts, chamber volume, and equil. time.
03
Press ▶ RUN
Watch all three sensor traces animate live.
04
Read the Result
PASS or FAIL with full tolerance breakdown.
PRO TIP: Use ⊕ t₂ FOCUS zoom while simulation is paused during equilibration for maximum detail.
THANK YOU
ThermoTest Pro — a complete BS EN 285 Annex C thermometric test simulator, running entirely in your browser. Built for sterilization engineers, validation specialists, and CSSD professionals.
© Palash Chandra Das  ·  pres.net.in  ·  BS EN 285 Thermometric Test Simulator